1. Check the possible causes and find out the solution!

     

    Possible Causes

    What You Can Do?

    Improper standard dilution

    Use appropriate diluent as blank. Make sure that the dilution is performed as according to protocol

    Standard improperly reconstituted

    Briefly spin standard vial before opening. Make sure that there is no undissolved material after reconstituting

    Standard degraded

    Store standards as according to protocol

    Curve doesn’t fit the scale

    Try plotting log-log or 5 parameter logistic curve fit

    Pipetting error

    Calibrate pipettes to make sure that the correct volume is dispensed

    Incomplete washing

    Increase washing cycles

  2. Check the possible causes and find out the solution!

     

    Possible Causes

    What You Can Do?

    Improper washing

    Make sure that the washing is done as according to protocol

    Poor mixing of samples

    Mix samples gently and evenly

    Dirty plate

    Make sure that the bottom of plate is clean

    Reagents too old

    Make sure that the reagents are not expired. Use freshly prepared reagents

    Bubbles in wells

    Make sure that there is no bubble in wells before reading

    Inconsistent pipetting

    Calibrate pipettes to make sure that the correct volume is dispensed

    Edge effects

    Make sure that the plate and reagents are equilibrated to room temperature before starting assay

  3. Check the possible causes and find out the solution!

     

    Possible Causes

    What You Can Do?

    Too much antibodies was used

    Reduce the concentration of primary or secondary antibodies

    Antibodies bind nonspecifically

    Use blocking buffer or choose another affinity-purified antibody

    Too much substrate reagent used

    Use substrate with higher dilution

    Insufficient washing

    Increase washing cycles

    Wrong concentration of blocking reagent

    Check the recommended concentration of blocking buffer

    Reaction not stopped

    Stop reactions with STOP buffer before reading

    Plate left too long before reading

    Take measurements shortly after addition of substrate and STOP buffer

    Insufficient Tween in wash buffer

    Use PBS+0.05% Tween as wash buffer

    Incubation temperature too high

    Optimize incubation temperature for each experiment

    Plate stacking during incubation lead to uneven temperature throughout the plate

    Avoid stacking plates together during incubation

    Pipetting error

    Calibrate pipettes to make sure that the correct volume is dispensed

    Reagents not mixed properly

    Make sure that all reagents are mixed properly and equilibrated to room temperature before assay

    Salt concentration of incubation and wash buffer

    Increase salt concentration to reduce nonspecific interaction

    Substrate incubation carried out in light

    Perform substrate incubation in dark

    Dirty plate

    Make sure that the bottom of plate is clean

  4. Check the possible causes and find out the solution!

     

    Possible Causes

    What You Can Do?

    Insufficient coating

    Use more antigens or antibodies for coating

    Substrate reagents have expired or prepared at a wrong pH

    Use fresh substrate reagents

  5. Check the possible causes and find out the solution!

     

    Possible Causes

    What You Can Do?

    Assay set up incorrectly

    Make sure that the instructions in the protocol is followed carefully

    Incorrect secondary antibody used

    Check if the correct secondary antibody is used

    Insufficient antibodies used

    Increase concentration of primary or secondary antibody

    Substrate reagents not fresh

    Use fresh substrate reagents

    Wrong settings of plate reader

    Check the settings (wavelength, filters, gain etc) of plate reader

    Insufficient incubation

    Follow the incubation time as indicated in the protocol booklet

    Sample concentration falls below detection limits of kit

    Decrease dilution factor or concentrate samples

    Plate washing too vigorous

    Check the setting of plate washer. Pipette wash buffer into wells gently

    Wells dried out

    Cover plate with adhesive film or incubate in humidified chamber throughout experiment

    Enzyme inhibitor present in buffers or reagents

    Inhibitors such as Sodium Azide can affect enzyme and assay performance. Ensure that there is no enzyme inhibitor in any buffers